COMMAND DETAILS
123
:SEQuence:END
Channel Specific
Command
Description
Sets the number of points in the sequence.
Syntax :SEQuence:END <NR1>
Parameter <NR1> Points
1~120 1~120.
Example :SEQ:END 5
Sets the max number of
points to 5.
Query Syntax :SEQuence:END?
Return Parameter <NR1> Points
1~120 1~120
Query Example :SEQ:END?
5
There are 5 points in the
sequence.
:SEQuence:POINt:RESistance
Channel Specific
Command
Description
Sets the resistance value for the current point. CR
mode only.
Syntax :SEQuence:POINt:RESistance <NRf>[OHM], MIN,
MAX
Parameter <NRf>[OHM], MIN, MAX Resistance value
100
100Ω
100 OHM
100Ω
MAX/MIN Maximum or minimum
value.
Example :SEQ:POIN:RES 100
Sets the resistance to 100.
Query Syntax :SEQuence:POINt::RESistance?
Return Parameter <NR1> Resistance Value
1 unit = 1 ohm Returns the resistance
value.
PEL-2000 Programming Manual
124
Query Example :SEQ:POIN:RES?
100
The resistance value is
100 ohm.
:SEQuence:POINt:CURRent
Channel Specific
Command
Description
Sets the current value for the current point. CC
mode only.
Syntax :SEQuence:POINt:CURRent <NRf>[A], MIN, MAX
Parameter <NRf>[A], MIN, MAX Current value
10 10A
100 A 100A
MAX/MIN Maximum or minimum
value.
Example :SEQ:POIN:CURR 1
Sets the current to 1.
Query Syntax :SEQuence:POINt::CURRent?
Return Parameter <NR1> Current Value
1 unit = 1 A Returns the current value.
Query Example :SEQ:POIN:CURR?
1
The current value is 1
amp.
:SEQuence:POINt:RISE/FALL
Channel Specific
Command
Description
Sets the rising and falling slew rates for the current
point.
Syntax :SEQuence:POINt:RISE/FALL<NRf>[A/us], MIN, MAX
Parameter <NRf>[A/us], MIN, MAX Slew rate
1.2 1.2A/us
1.2 A/us 1.2A/us
MAX/MIN Maximum or minimum
value.