Real-time functions and outputs
QuantumX A03031_24_E00_01 HBM: public 163
S Filters set for the MX878B/879B do not apply to RMS channels. These
channels are always unfiltered. However, the input signal is filtered.
Peak value function
Keep in mind the following notes when working with peak value monitoring
channels:
S The maximum sampling rate is 4800 Hz
S The sampling rate of the channel for peak value monitoring must not be
higher than the sampling rate of the input channel.
S The peak value units do not allow for any other peak value units or RMS
values as input
Each peak value detection unit can monitor either the Min peak or the Max
peak of one of the four signals within the system identified as "isochronous".
A peak value unit may operate in different modes:
S EXPORT: peak values will be continuously updated
S HOLD: last peak value will be "frozen"
S PEAK VALUE: peak value detection enabled
S FOLLOW: peak value detection disabled, i.e. the channel returns the origi
nal signal of the input channel
The following combinations are possible:
RUN MAX PEAK VALUE
HOLD MAX PEAK VALUE
FOLLOW HOLD MAX
This also applies to minima.