148 A4774-1.0 HBM: public WTX110
from the filled sack, so that the released sack only weighs 49.8 kg instead of
50 kg.
In this case enter 20 as the correction factor for Syst.diff., resulting in overfilling
of 200 g (10 * P1 of SYD). As a result, the sack will initially be filled with
50.2 kg, and after release will have the target weight of 50 kg.
10.2.24 Maximum dosing time (Max.dos.t(s))
The dosing time starts when dosing starts (RUN command or digital input or
Enter key) and ends with the checkweighing (Finished signal). If the maximum
dosing time has been exceeded, the dosing process is aborted with an error,
and the coarse and fine flow outputs are deactivated (see section 10.2.10
"Alarm (Alarm:Empty)").
10.2.25 Tare delay (Tare delay(s))
You can use this time for example to blank out interference from putting up
sacks or putting on containers. Set the tare delay so that any such
disturbances have been eliminated before taring begins.
When the filling process starts (RUN command or digital input or Enter key) the
tare delay starts. Taring is then carried out when the tare delay ends.
Requirements
In upward dosing, the empty weight and coarse flow cut-off point must not be
exceeded on starting.
Tare mode must be active (Tare Mode On).
Activation
Activate taring after a delay by way of the tare delay, specifying the delay time
to apply (see also section 10.2.14 "Empty weight (Empty weight)")