WTX110 A4774-1.0 HBM: public 209
16.2 Error messages
In the event of an error during calibration or in weighing operation, the following
messages may be displayed:
Error message display Possible causes Remedy
ADM faulty A/D converter not
supplying data
Replace A/D converter
Short-circuit in load cell
cable
Check wiring
EMC interference Check wiring
ADM not factory cal.
Invalid Setup data
Calibration faulty Repeat calibration
ADM faulty Replace ADM
ADM not installed ADM not installed
ADM not correctly
inserted
Check whether the A/D
converter is installed
Insert ADM (correctly)
Calibration locked
Error Calibr. Jumper
Jumper for calibration
lock in locked position
Remove bridge
I/O error
Interface error Check interface and
replace if necessary
Incline sensor Optical tilt sensor reporting error
Inclination too great Change inclination
Optical tilt sensor faulty Replace optical tilt sensor