Calibration
188 A4774-1.0 HBM: public WTX110
The following table lists default values of the Calibration group to which
Reset Parameter resets.
Group Parameters Standard
value
Your calibration
1 Scale parameters Single/dual/triple
range
Single
range
One/two/three
intervals
One interval
Capacity 3000
Interval 1
Unit kg
2 Calibration Geo value 20
Zero (mV/V) W1 0.00000
Load (mV/V) W1 2.00000
Zero (mV/V) W2 0.00000
Load (mV/V) W2 2.00000
5 Adaptation Motion window 0.5D
Motion count(er) 20
Filter size 11
Auto zero range 0.5D
Pushbutton zero (+) 2%
Pushbutton zero (–) 2%
Power-up Zero OFF
Overload 9D
Incline switch OFF
NTEP N
Underload 20d Y
With taring Y
Update rate 225