Calibration
190 A4774-1.0 HBM: public WTX110
Function/Secondary selectionKeyMenu
Y Calibrate zero point
N Continue to step Zero(mV/V)
Calibrating... Measure the zero point of the scale
Zero: 0.00 Display the 10 times higher weight value
resolution (as a check)
Zero(mV/V): 0.40000 Display the standardized signal
(20 % dead load=20 % signal)
LC capacity 1000 Enter the maximum capacity of a load cell
in the unit selected in group 1 (Scale
Parameters\Unit)
No. of LCs 3 Enter the number of connected load cells: max. 8
mV/V of LC1 1.99995 Enter the sensitivity of the first load cell.
Continue with the next load cell. After the last
load cell continue with the next step.
Load (mV/V): 1.40000 Display the standardized signal for the maximum
load (20 % dead load + 50 % net load)
Information
The Clr key can be used to clear the value and enter a new one.
Information
To enter decimal values press the F2 key 10 times. A decimal point appears in
the display.