Operating Precautions
5
Follow these precautions to ensure safe operation and to obtain the full benefits of the various functions.
Operating Precautions
Before Use
If the connection cable or the instrument is damaged, there is a risk of electric shock.
Before using the instrument, perform the following inspection.
Before using the instrument, check that the coating of the connection cables are nei-
ther ripped nor torn and that no metal parts are exposed. Using the instrument under
such conditions could result in electrocution or a short-circuit. Replace the connec-
tion cables with those specified by our company.
Before using the instrument for the first time, verify that it operates normally to
ensure that no damage occurred during storage or shipping. If you find any damage,
contact your authorized Hioki distributor or reseller.
To measure accurately
Warm up the instrument for more than 30 minutes before use.
To maintain sufficient measurement accuracy of the instrument, be sure to help heat radiation.
Example: Keeping away the instrument from a heat, leaving sufficient clearances around the instrument,
installing cooling fans to the rack in which the instrument is mounted, or other measures.