3.2 Configuring Settings
48
This section describes how to set the synchronization source used to determine the cycle (between zero-
cross events) that will be used as the basis for calculations.
You can select from the following parameters: U, I, DC (fixed at 200 ms)
3.2.4 Setting the Synchronization Source (SYNC)
When the synchronization source is set to I or DC,
the SYNC lamp will light up. When set to the volt-
age U, the SYNC lamp will not light up.
1 Press .
2 Press to switch to the synchroni-
zation source settings screen. ([SynC.]
will be displayed in display area
a.)
3 Press or until the display area
a flashes.
4 Press or to set the
synchronization source.
Default setting: U
Settings:
U i dC U ⋅⋅⋅
(Setting the source to I or DC will cause the SYNC
lamp to light up.)
5 Press to exit the settings.
The instrument will return to the normal display (Mea-
surement screen).