3.7 Using D/A Output
94
Example uses
D/A output can be used in combination with a data logger or recorder. For more information about the output
voltage, output resistance, response time, and output update rate, see "Chapter 5 Specifications" (p.135).
When using a VT ratio or CT ratio, output voltages are determined by multiplying the range value by the
VT ratio or CT ratio.
When the measurement range is set to auto-range operation, the output rate for level output and high-
speed level output will also change with changes in the range. When measuring lines characterized by
abrupt changes in measured values, exercise care not to mistake the range conversion. It is recom-
mended to use a fixed range in measurement applications such as this.
Output of 0 V is generated while displaying invalid data.
When using a VT ratio or CT ratio, the value obtained by multiplying the range by the VT ratio or CT ratio
serves as the 1 V RMS value.
When the measurement range is set to auto-range operation, the output rate for waveform output will also
change with changes in the range. When measuring lines characterized by abrupt changes in measured
values, exercise care not to mistake the range conversion. It is recommended to use a fixed range in mea-
surement applications such as this.
Waveform output will change even in the display hold state and during averaging processing.
Example uses of level output
Example uses of waveform output