3.10 When warning lamp, o.r, or the Unit Indicator Flashes
114
3.10.4 When the Unit Indicator Flashes
Status Solution
The unit indicator flashes while
displaying the average
The average value being displayed contains [o.r] data. If no [o.r] data is present during
averaging processing, the unit indicator will not flash. When [o.r] is displayed, internal
data for which the voltage or current exceeded 152% of the range, or the active power
exceeded 231.04% of the range, will be used as-is in calculating the average value.
The integrated value or time aver-
age value unit indicator flashes
(TOTAL and T.AV also flash)
Reset integrated values, change the range, and repeat integration. If no PEAK OVER
condition occurs during integration, the indicator will not flash.
See: "3.3 Integration" (p.61)