3.9 Other Functions
110
This section describes how to initialize the instrument’s settings. Initializing the instrument (performing a sys-
tem reset) reverts settings to their default values at the time the device was shipped from the factory. The sys-
tem reset should be performed while the self-test is in progress after turning on the instrument (before the
display changes to the normal display).
Example: PW3335-03
3.9.4 Initializing the Instrument (System Reset)
Disconnect any voltage or current input from the instrument before performing the reset.
The RS-232C communications speed, GP-IB address, and LAN-related settings will not be initialized.
1 Turn on the instrument.
2 While the product model and ver-
sion are being displayed, press
and then .
(The SHIFT lamp will not light up.)
The System Reset screen will be displayed,
and settings will be reverted to their default val-
ues when the instrument was shipped from the
factory.