3.2 Configuring Settings
39
1 Press .
2 Press to switch to the current
input settings screen. ([SEnSor] will be
displayed in display area
c.)
3 Press or until the display area
d flashes.
4 Press or to set the current
input method.
Default setting: OFF
Settings: oFF (direct input) tyPE.1 tyPE.2 ⋅⋅⋅
(When set to TYPE.1 or TYPE.2, the SENSOR lamp
will light up.)
The current input method cannot be changed while integration is being performed or during display hold or
maximum value/ minimum value display hold operation.
Auto-range integration mode cannot be used when the current input method is set to TYPE.1 or TYPE.2.
5 Press to exit the settings.
The instrument will return to the normal display (Mea-
surement screen).