3.2 Configuring Settings
51
Display of frequency measurements (Displayed as V Hz or A Hz on panel display)
2 Press to switch to the zero-cross
filter settings screen. ([Fr.] will be dis-
played in display area
b.)
3 Press or until the display area
b flashes.
4 Press or to set the zero-
cross filter (100 Hz/ 500 Hz/ 5 kHz/
100 kHz).
Default setting: 500 Hz
Settings: 500 5k 100k 100 500 ⋅⋅⋅
5 Press to exit the settings.
The instrument will return to the normal display (Mea-
surement screen).
Because measured values become difficult to read near the point at which the display switches, the resolu-
tion is lowered by one digit.
Example: If the frequency being measured changes from 1 kHz to 999 Hz, the display resolution for 1 kHz to
9.9999 kHz measurement will remain in effect, and the value will be displayed as 0.9990 kHz. If
the frequency being measured then changes again to a value that is less than or equal to 990 Hz,
the instrument will switch to the display resolution for 100 Hz to 999.99 Hz and display the value
as 990.00 Hz.
See: Frequency Measurement Specifications (p.142)
From 0.1000 Hz to 9.9999 Hz From 10 Hz to 99.999 Hz
From 100 Hz to 999.99 Hz From 1 kHz to 9.9999 kHz
From 10 kHz to 99.999 kHz
100 kHz