Chapter 9. The Qualification Study
Kaye Validator AVS User’s Manual 99
9.3 Viewing the Active Qualification Study (cont.)
• Open, under-range, over-range or failed sensors are not allowed to participate in
calculations or events.
• Each box lists the SIM number, group name, sensor label and associated measurand
(temperature, humidity, pressure, etc.) If there are more sensors or calculations than fit on
the screen, use the scroll bar to display the remaining data.
• A button to stop qualification is available at the bottom of the screen. Data is refreshed
as per display rate 5 or 10 seconds selected on preferences screen the rates shown below:
Note: After disconnecting the console the Validator AVS continues the study. Live monitoring
data are discarded after disconnecting and will resume on reconnecting the console.
Depending on the setup the exposure can be started and stopped manually or will start automatically
depending on the setup definition (in that case that “Start exposure” Button is not available).
To stop the study, press the Stop Qual button.
After stopping the study the Save Study button becomes available. Please save the study to
your console. If a console is connected to a Kaye Validator AVS with an unread study it will
insist on reading the study before starting any new study.