Chapter 6. Defining Study Setups
Kaye Validator AVS User’s Manual 72
6.7 Specifying Group Calculations
Figure 43: Calculations
Once you have defined sensors and assigned them to one or more groups, the next step is to
specify the calculations to be performed on the group of sensors during the qualification
study. Calculations include statistical, lethality, saturation pressure of steam and saturation
temperature of steam calculations.
The Statistical Calculations are all fixed preselected by default and will be all calculated.
Note: Lethality calculations needs to be defined in the setup!
It is also possible to define events to monitor during the study. These events are listed in the
reports and can be imported into the define cycles screen of the reporting tool.
When finished, press Calibration Parameters to enter the qualification parameters.