Chapter 10. Generating Reports
Kaye Validator AVS User’s Manual 103
10.2 Report Types
10.2.1 Setup Report
The Setup Report documents sensor labels and definitions for each SIM, wiring locations, and the
parameters for sensor calibration, qualification and system preferences. Individual group
information includes the sensors assigned to that group, statistical, lethality, and saturated steam
calculations, the event defined for that group, and group-specific information such as vessel ID,
SOP/protocol number, load description, asset details (if associated with a report) and comments.
Sensor calibration parameters can include high, low and check set point, Stability and Deviation
criteria defined while creating the setup. Qualification parameters include Start stop conditions,
scan rate and data storage. For current, voltage and pressure sensors, the report lists scaling
information such as min and max of input and output. System preferences such as temperature
units, pressure units, and line frequency also appear. The report header lists the date and time the
report was printed, the setup name, your company name, the firmware and software version
numbers, the date this setup was saved, and identifies the person who saved this setup.
The Setup Report can be generated directly from the setup review page or using the reporting tool.
• When you generate this report independent of the full study, you can use the Setup Report to
check your setup configuration and/or as a guide for wiring SIMs.