Chapter 10. Generating Reports
Kaye Validator AVS User’s Manual 106
10.2.3b
Qualification Summary
R
e
p
o
r
t
The Qualification Summary Report summarizes your study. You can apply a number of
calculations, and generate a variety of calculation graphs.
The report header lists the date and time the report was printed, the setup name, your company
name, the name of the person who performed the qualification study, the run number, the Kaye
Validator AVS serial number, and the firmware and software version numbers.
The overview section lists the groups included in the report, the date and time the study started
and ended, the duration of the qualification study, the duration of the exposure cycle, the SIMs
used in the study by slot number and serial number, and user comments entered during
post-qualification reporting.
For each group, the Summary Report documents interval calculations, lethality calculations,
any sensors that have been deleted from group calculations, event messages generated
during the qualification study, and group events that were added during post-qualification
reporting.
• For lethality calculations, the Summary Report documents the lethality calculation
parameters, the lethality calculation condition, the lethality calculation start/stop times, and
the minimum lethality for that group by sensor label, value, and location.
• If you enabled the Lethality option on the Print Options screen, the total accumulated
lethality for each sensor is reported.
10.2.4 Calibration Verification Report
The Calibration Verification Report documents the SIMs used in calibration verification by
slot number and serial number, the calibration setpoints, stability and calibrated deviation
criteria, and the evaluation of stability, deviation, and the results for each temperature
setpoint.
Note: If a sensor fails calibration verification, it is marked with a double asterisk (**)
to the right of the value in the temperature column, and from then on it is marked as Failed.
The report header lists the date and time the report was printed, and identifies the name of the
setup used to verify sensor calibration, your company name, the Kaye Validator AVS serial
number, and the firmware and software version numbers.