Chapter 10. Generating Reports
Kaye Validator AVS User’s Manual 105
10.2.3 Qualification Reports (cont.)
The following sections of the Qualification Report are documented on a group basis. Data is
reported at the data storage rate specified in the setup or the print sampling rate defined in
post-qualification reporting. When an event occurs, whether it is Start Qualification, Start
Exposure, Stop Exposure, Stop Qualification, or one of the group events, the message appears
in each section of the report with a snapshot of the data at that time. If this was a manual event,
the name of the person who manually started or stopped the event is also printed. Data is
displayed in columns with shaded areas to help you locate data associated with a specific time
period. Sensors that are excluded from post-qualification calculations are marked with an
asterisk (*).
The Sensor Inputs section documents the raw data collected during the study.
If you are conducting penetration studies, the Lethality Calculations section documents
accumulated lethality for each sensor, and the minimum accumulated lethality value and its
location.
The Statistical Calculations section documents the calculations specified in the setup and/or
the calculations specified during post-qualification reporting. This section also documents
exposure time, cycle time, and temperature standard readings.
The Saturated Steam Calculations section documents the saturation temperature of steam
calculation and the saturation pressure of steam calculation.
The last section of the report, System Messages, consists of two parts. The first part
describes events that occurred during the qualification study.
These are the events that are common to all groups:
Start/Stop Qualification
Start/Stop Exposure
The second part of this section documents event timer intervals added during post-
qualification reporting.