Chapter 7. Calibrating and Verifying Sensors
Kaye Validator AVS User’s Manual 85
Preview stability criteria
Calibration Checklist -- A calibration checklist allows users to keep track of their
particular step in the calibration:
Temperature Reference Set
Sensors Stabilized
IRTD Temperature standard stabilized
Deviation calculated on uncalibrated sensors
Sensors Calibrated
Deviation calculated on calibrated sensors
Saving Calibrated Temperature to files
Calibration point complete
A green check mark appears next to each step in the process when it is completed.
7.5.1 Reasons for Calibration Failure
Calibration or verification might fail for one of four reasons:
• Failed calibrated deviation
• Failed uncalibrated deviation
• Failed calibration due to Unstable IRTD and loggers
• Failed calibration due to not enough data
Data for a live calibration or verification refreshes every 5 or 10 seconds as per display rate.