Chapter 9. The Qualification Study
Kaye Validator AVS User’s Manual 96
9.2 Load a Setup
Before the start of the qualification process, make sure to load the defined setup for this asset from
the Console into the Kaye Validator AVS.:
• Select the predefined setup file for this asset
• Load the setup into the Kaye Validator AVS
IMPORTANT: When loading a setup with a different calibration signature (calibration
parameters and/or temperature units) than the setup currently in memory,
any sensors already calibrated will be marked as uncalibrated. If you reload
the original setup, and you have not calibrated sensors using the new setup,
the offsets are restored and the sensors are marked as calibrated.
When loading a setup with a different line frequency than the setup currently in memory, the Kaye
Validator AVS must stabilize for approximately 2 - 3 minutes before running a qualification study
or invalid data is displayed.
9.2.1 Select a Setup File
If the setup is not yet created please see chapter 6 for instruction how to create a setup.
If a setup for the asset is available, go to the Asset Details page and, from the Setup pane
select the desired set up and press Initiate Qualification.
The pop window gives the opportunity to change the SOP Protocol Number and display
the run number. The run number is counting up automatically but is changeable if
required. Pressing Ok opens the “Select AVS” screen.