Chapter 10. Generating Reports
Kaye Validator AVS User’s Manual 104
10.2.2 Calibration Report
The Calibration Report documents the SIMs used in calibration by slot number and serial
number, the calibration setpoints, stability and deviation criteria, and the evaluation of
stability, deviation, and the results for each temperature setpoint.
Note: If a sensor fails calibration, it is marked with a double asterisk (**) to the right of the
value in the temperature column, and from then on it is marked as Failed.
The report header lists the date and time the report was printed, and identifies the name of the
setup used to calibrate the sensors, your company name, the Kaye Validator AVS serial number,
and the firmware and software version numbers.
10.2.3 Qualification Reports
There are two different reports that document the qualification study: the Qualification
Detailed Report, and the Qualification Summary Report.
10.2.3a
Qualification Detailed
R
e
por
t
The Qualification Detailed Report organizes information for each group by sensor data and
type of calculation (lethality, statistical, saturated steam). You can customize the
Qualification Report by specifying the groups to be reported, add events based on the stored
data, and add explanatory notes and comments to the beginning or end of the report.
The report header lists the date and time the report was printed, the name of the setup the
qualification study is based on, the name of the person who performed the study, your
company name, the Kaye Validator AVS serial number, the qualification study run number,
and the firmware and software version numbers.
The overview section documents the groups included in the report, any sensors excluded from
group calculations, the SIMs used in the study by slot number and serial number, and user
comments added during post-qualification reporting.