Chapter 10. Generating Reports
Kaye Validator AVS User’s Manual 126
Avg
-Min
Lethality
Sat P
Sat T
• MKT calculations
Text boxes enable you to select the minimum and maximum values of the Y axis for the
selected calculation graph. (These boxes accept only numerics, along with the minus
sign, decimal point and comma.)
Above the Groups/Sensors pane, three tabs allow you to select the Lethality, Saturation
Pressure and Saturation Temperature calculations.
The Lethality calculation is available only if temperature sensors are present in the study.
The left pane has the lethality graph for the entire study, while the right pane enables you to
change the D, Z and Base temperature values.
The Saturation Pressure calculation is available only if the study includes at least one
temperature sensor along with pressure sensors.
The Saturation Temperature calculation is available only if the study includes at least one
pressure sensor along with temperature sensors.
No calculations are available for a contact sensor group.
Based on the group selection, the following calculation graphs are generated:
Temperature Sensors
Temperature Sensors Calculations
RH (Humidity) Sensors
RH Sensors Calculations
Pressure Sensors