3-18 Basic DMM Operation Model 2701 User’s Manual
Model 7700 switching module
Connections for the Model 7700 switching module are shown in Figure 3-7. Note that
only channels 21 and 22 can be used for current measurements.
Figure 3-7
DCI and ACI connections using Model 7700 switching module
Amps measurement procedure
NOTE Make sure the INPUTS switch is in the correct position. To use front panel
inputs, it has to be in the “F” (out) position. For switching modules, it has to be
in the “R” (in) position.
1. If a switching channel is presently closed (displayed), press OPEN to open it.
2. Select the amps measurement function by pressing DCI or ACI.
3. Use the RANGE and keys to select a measurement range consistent with the
expected current or press AUTO to select autoranging (AUTO annunciator turns
on). Details on range are provided in Section 4.
4. Apply the current(s) to be measured.
CAUTION Do not apply more than 3A to the input or the AMPS fuse will blow.
Model 7700 switching module — When performing current measure-
ments, the maximum allowable voltage is 60VDC or 30V RMS.
Exceeding these limits could cause damage to the switching module.
5. If using a switching module, use the
and keys to close the desired amps
channel (for the Model 7700, 21 or 22). All other channels will be open.
Model 7700
Switching
Module
H
L
CH 21 or 22
Caution: Maximum input: 60VDC or 30V RMS, 3A switched,
60W, 125VA maximum.
Current
Source

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals