B-6 Model 7700 Connection Guide Model 2701 User’s Manual
Wiring procedure
Use the following procedure to wire the Model 7700 module. Make all connections using
correct wire size (up to 20 AWG). Also, make sure to add supplementary insulation around
the harness for voltages above 42V peak (Figure B-4).
WARNING All wiring and supplementary insulation must be rated for the
maximum voltage in the system. For example, if 1000V is applied to
the front terminals of the DMM, the plug-in module wiring must be
rated for 1000V.
1. Make sure all power is discharged from the Model 7700 module.
2. Access the screw terminals (Figure B-2).
3. Using a small flat-blade screwdriver, loosen terminal screws and install wires as
desired. (Figure B-4 shows connections to channels 1 and 2.)
4. Route wire along wire-path and secure with cable tie as shown.
5. Fill in a copy of the connection log (Table B-1) and affix it to the module cover.
6. Close and lock cover.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm