5-20 Rel, Math, Ratio, Channel Average, dB Model 2701 User’s Manual
dB
Expressing DC or AC voltage in dB makes it possible to compress a large range of
measurements into a much smaller scope. The relationship between dB and voltage is
defined by the following equation:
where: V
IN
is the DC or AC input signal.
V
REF
is the specified voltage reference level.
The instrument will read 0dB when the reference voltage level is applied to the input.
If a relative value is in effect when dB is selected, the value is converted to dB then REL is
applied to dB. If REL is applied after dB has been selected, dB has REL applied to it.
NOTE The dB calculation takes the absolute value of the ratio V
IN
/ V
REF
.
The largest negative value of dB is -160dB. This will accommodate a ratio of
V
IN
= 1µV and V
REF
= 1000V.
dB conguration
Remote programming must be used to configure the Model 2701 for dB measurements. It
cannot be configured from the front panel.
Scanning
Typically a scan using dB is configured and run using remote programming. However,
once dB is selected using remote programming, a simple dB scan can be configured and
run from the front panel. When the simple scan is configured, it will use the dB
measurement setup for each channel in the scan. Details on configuring and running a scan
are provided in Section 7.
NOTE See Section 7 to configure and run a scan.
Remote programming dB
dB commands
The dB commands are listed in Table 5-4. Details on these commands follow the table.
dB 20log
V
IN
V
REF
-------------
=

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm