2-2 Close/Open Switching Module Channels Model 2701 User’s Manual
Close/open overview
NOTE For remote operations, you can use the internal web page of the Model 2701 to
send commands, queries, and take readings. See “Internal web page” in
Section 10 for details.
Also, supplied TestPoint Runtime start-up software can be used to control the
Model 2701. See Section 4 in the Model 2701 Instrument Networking
Instruction Manual for details.
NOTE This section covers basic close/open operations for switching module channels.
It also covers the operating characteristics that are unique to the Model 7700
switching module.
There are two modes of close/open operation:
System channel operationThis is the mode of operation that should be used
exclusively by most (if not all) users. When you close an input channel (or channel-
pair), other channels on the switching module close automatically to internally
connect it the DMM of the Model 2701.
Multiple channel operation — This mode of operation provides additional flexibility
by providing individual control of each switching module channel. However, careless
operation could create a safety hazard and/or damage the switching module and other
equipment. Multiple channel operation should only be used by experienced test
engineers.
CAUTION To prevent damage to a switching module, do not exceed the maximum
signal level input for that module. Most switching modules are rated
for 303V. The following command queries maximum module voltage:
SYSTem:CARDx:VMAX? ‘Request maximum allowable voltage for
‘CARDx (where x is the slot number for
‘the module).
For system channel operation, the instrument will display the
“OVERFLOW” message when the maximum allowable voltage for the
module is being exceeded.
However, for multiple channel operation, the “OVERFLOW” message
will not occur until the maximum voltage of the mainframe (not
module) is exceeded. Therefore, the “OVERFLOW” message would
occur only if 1010V is exceed.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals