8-14 Triggering Model 2701 User’s Manual
Remote programming triggering
Trigger model (remote operation)
The following paragraphs describe how the Model 2701 operates for remote operation.
The flow chart in Figure 8-10 summarizes operation over the bus. The flow chart is called
the trigger model because operation is controlled by SCPI commands from the Trigger
subsystem. Key SCPI commands are included in the trigger model.
Idle and initiate
The instrument is considered to be in the idle state whenever operation is at the top of the
trigger model. As shown in Figure 8-10, initiation needs to be satisfied to take the
instrument out of idle. While in the idle state, the instrument cannot perform any measure
or step/scan operations.
The following commands will return operation to the top of the trigger model (idle) at the
START point of the trigger model:
ABORt
*RCL 0, 1, 2, or 3
SYSTem:PREset
*RST
What happens next depends on the state of initiation. If continuous initiation is already
enabled, the instrument will leave the idle state. SYSTem:PRESet enables continuous
initiation. Therefore, operation will immediately leave the idle state when it is sent. The
*RCL command will do the same if INITiation:CONTinuous ON is a user-saved default.
*RST disables continuous initiation. Therefore, the instrument will remain in the idle
state.
Either of the following two initiate commands will take the instrument out of the idle state:
INITiate
INITiate:CONTinuous ON
NOTE While in remote, pressing the LOCAL key restores continuous front panel
operation.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals