13-4 SCPI Signal Oriented Commands Model 2701 User’s Manual
CONFigure:<function> [<rang>], [<res>], [<clist>]
CONFigure:VOLTage[:DC] [<rang>], [<res>], [<clist>] Configure DCV
CONFigure:VOLTage:AC [<rang>], [<res>], [<clist>] Configure ACV
CONFigure:CURRent[:DC] [<rang>], [<res>], [<clist>] Configure DCI
CONFigure:CURRent:AC [<rang>], [<res>], [<clist>] Configure ACI
CONFigure:RESistance [<rang>], [<res>], [<clist>] Configure Ω2
CONFigure:FRESistance [<rang>], [<res>], [<clist>] Configure Ω4
CONFigure:FREQuency [<rang>], [<res>], [<clist>] Configure FREQ
CONFigure:PERiod [<rang>], [<res>], [<clist>] Configure PERIOD
CONFigure:TEMPerature [<clist>] Configure TEMP
CONFigure:CONTinuity [<clist>] Configure CONT
Parameters <rang> = Range parameter for the specified function. For example, for
DCV, range parameter value 10 selects the 10V range. See
the “NOTES” that follow Table 13-1 for additional information.
<res> = 0.1 i.e., 100.0 V (3½ digits)
0.01 i.e., 10.00 V (3½ digits)
0.001 i.e., 1.000 V (3½ digits)
0.0001 i.e., 1.0000 V (4½ digits)
0.00001 i.e., 1.00000 V (5½ digits)
0.000001 i.e., 1.000000 V (6½ digits)
The resolution of the <res> parameter value and the selected
range sets the number of display digits. As shown above, with
the 100V range selected and <res> = 0.1, a 100V reading will
be displayed as 100.0 V (3½ digits).
The display will default to 3½ digits when using parameter
values that attempt to set the display below 3½ digits. For
example, a 10V reading using <res> = 0.1 for the 10V range is
displayed as 10.00 V, not 10.0 V.
A command using parameter values that attempt to set the
display above 7½ digits is ignored and generates error -221
(settings conflict error).
The <res> parameter is ignored when a <clist> is included in
the command string. Resolution for the scanlist channel(s) is
determined by the present setting for the specified function and
by the present resolution setting for the specified function.
See the “NOTES” that follow Table 13-1 for additional
information.
<clist> = Channel(s) in the scanlist to be configured. When the channel
list parameter (<clist>) is included, the present instrument
settings are not affected. Instead, the channel(s) in the <clist>
for the specified function is configured. See the “NOTES” that
follow Table 13-1 for additional information.