Model 2701 User’s Manual Specifications A-7
Accuracy calculations
The information below discusses how to calculate accuracy for both DC and AC
characteristics.
Calculating DC characteristics accuracy
DC characteristics accuracy is calculated as follows:
Accuracy = ±(ppm of reading + ppm of range)
(ppm = parts per million and 10ppm = 0.001%)
As an example of how to calculate the actual reading limits, assume that you are
measuring 5V on the 10V range. You can compute the reading limit range from one-year
DCV accuracy specifications as follows:
Accuracy = ±(30ppm of reading + 5ppm of range)
±[(30ppm
×
5V) + (5ppm
×
10V)]
±(150µV + 50
µ
V)
±200µV
Thus, the actual reading range is: 5V± 200µV or from 4.9998V to 5.0002V.
DC current and resistance calculations are performed in exactly the same manner using
the pertinent specifications, ranges, and input signal values.
Calculating AC characteristics accuracy
AC characteristics accuracy is calculated similarly, except that AC specifications are given
as follows:
Accuracy = (% of reading + % of range)
As an example of how to calculate the actual reading limits, assume that you are
measuring 120V, 60Hz on the 750V range. You can compute the reading limit range from
ACV one-year accuracy specifications as follows:
Accuracy = ±(0.06% of reading + 0.03% of range)
±[(0.0006
×
120V) + (0.0003
×
750V)]
±(0.072V + 0.225V)
±0.297V
In this case, the actual reading range is: 120V± 0.297V or from 119.703V to 120.297V.
AC current calculations are performed in exactly the same manner using the pertinent
specifications, ranges, and input signal values.
Appendix A.fm Page 7 Tuesday, July 22, 2003 11:10 PM

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals