Model 2701 User’s Manual Signal Processing Sequence and Data Flow D-9
[SENS[1]]:DATA[LATest]?
[SENS[1]]:DATA:FRESh?
These commands are used to return (read) the last processed data array stored in the
sample buffer.
[SENS[1]]:DATA[:LATest]?
This command returns (reads) one data array. It returns the last processed data array stored
in the sample buffer. If, for example, 10 data arrays are stored in the sample buffer, only
the last (10
th
) data array is returned.
DATA? does not affect data in the sample buffer. Therefore, subsequent executions of
DATA? acquires the same data array. In order to return a new reading, you must first
trigger a new reading(s) and then use DATA?.
When using DATA? to retrieve data, it is good practice to include reading numbers in the
data arrays. Reading numbers that do not change will indicate that the same data array is
being returned.
NOTE The FORMat:ELEMents command (see Section 14) is used to include the
reading number (RNUM) element in each data array.
[SENS[1]]:DATA:FRESh?
This command is similar to DATA[:LATest]? in that is also returns the last processed data
array stored in the buffer. However, it can only be used once to retrieve the same data
array. That is, the data array reading must be “fresh.” Sending this command again to read
the same (stale) data array will not work. It will generate an error (-230; data corrupt or
stale) or cause the Model 2701 to time-out. In order for DATA:FRESh? to respond, you
must first trigger a new (fresh) reading.
Using this command to retrieve data ensures that only new (fresh) readings are returned
(no readings are repeated). However, as previously noted, problems may occur if a new
reading is not ready (available) when this command is executed.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals