Model 2701 User’s Manual Basic DMM Operation 3-23
Cable leakage
For high resistance measurements in a high humidity environment, use Teflon™ insulated
cables to minimize errors due to cable leakage.
Standard resistance measurements
NOTE Make sure the INPUTS switch is in the correct position. To use front panel
inputs, it must be in the “F” (out) position. For switching modules, it must be in
the “R” (in) position.
Perform the following steps to measure resistance:
1. If a switching channel is presently closed (displayed), press OPEN to open it.
2. Select the ohms measurement function by pressing 2 or 4.
3. Use the RANGE
and keys to select a measurement range consistent with the
expected resistance or press AUTO to select autoranging (AUTO annunciator turns
on). Details on range are provided in Section 4.
4. Connect the resistance(s) to be measured.
CAUTION Front panel inputs — Do not apply more than 1000V peak between
INPUT HI and LO or instrument damage may occur.
Model 7700 switching module — Do not apply more than 300V DC or
300V RMS (425V peak) for AC waveforms between input high (H) or
input low (L) or switching module damage may occur.
5. If using a switching module, perform the following steps to close the desired
channel. Keep in mind that for 4 measurements, you will close the primary
(INPUT) channel (1 through 10). The paired channel will close automatically.
a. Press the CLOSE key.
b. Use
, , , and to key in the channel number and press ENTER. The
previously closed channel(s) (if any) will open, and the specified channel (or
channel pair) will close.
NOTE While in the normal measurement state, you can use the and keys to close
channels. In general, each key press will open the presently closed channel and
then close the next higher or lower channel.
6. Observe the displayed reading. If the “OVERFLOW” message is displayed, select
a higher range until a normal reading is displayed (or press AUTO for
autoranging). For manual ranging, use the lowest possible range for the best
resolution.
7. To measure other switching channels, repeat steps 5 and 6.
8. When finished, press OPEN if there is a channel closed.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm