A-10 Specifications Model 2701 User’s Manual
Optimizing measurement accuracy
The configurations listed below assume that the multimeter has had factory setups
restored.
NOTE
For maximum system performance, it is recommended that all measurement
cables be limited to less than 3 meters.
DC voltage, DC current, and resistance:
• Select 6
½
digits, 10 PLC, filter ON (up to 100 readings), fixed range.
• Use REL on DC voltage and 2-wire resistance measurements.
• Use 4-wire resistance measurements for best accuracy.
AC voltage and AC current:
• Select 6
½
digits, 10 PLC, filter ON (up to 100 readings), fixed range.
Temperature:
• Select 6
½
digits, 10 PLC, filter ON (up to 100 readings).
Optimizing measurement speed
The configurations listed below assume that the multimeter has had factory setups
restored.
DC voltage, DC current, and resistance:
• Select 3
½
digits, 0.002 PLC, filter OFF, fixed range.
AC voltage and AC current:
• Select 3
½
digits, 0.002 PLC, filter OFF, fixed range.
Temperature:
• Select 3
½
digits, 0.002 PLC, filter OFF.
For all functions, turn off the display, autozero, and set the trigger delay to zero. Use the
:SAMPle:COUNt and READ? bus commands.