EasyManuals Logo
Home>Keithley>Multimeter>2701

Keithley 2701 User Manual

Keithley 2701
506 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #228 background imageLoading...
Page #228 background image
7-2 Scanning Model 2701 User’s Manual
Scanning fundamentals
The Model 2701 can scan the channels of up to two installed Keithley switching modules.
Each scan channel can have its own unique setup. Aspects of operation that can be
uniquely set for each channel include function, range, rate, AC bandwidth, rel, filter,
digits, math, offset compensation, TEMP transducers, limits, channel average, channel
ratio, and volts dB.
NOTE Readings for scanned channels are automatically stored in the buffer. With
buffer auto clear enabled (which is the default), the buffer clears when the scan
is started. When disabled, scan readings are appended to the buffer. The
TRACe:CLEar:AUTO <b> command is used to enable/disable buffer auto clear
(Section 6).
A pseudocard can be installed in an empty slot. With the 7700 pseudocard
installed, the instrument will operate as if a Model 7700 switching module is
installed in the slot. This allows the user to configure a scan and exercise its
operation before the switching module is installed in the Model 2701. Use the
following commands to install 7700 pseudocards in empty slots:
SYSTem:PCARd1 C7700 ‘Install’ 7700 pseudocard in slot 1.
SYSTem:PCARd2 C7700 ‘Install’ 7700 pseudocard in slot 2.
Pseudocards for other switching modules can instead be installed. Details on
installing other pseudocards are provided in Section 2.
There are SCPI commands to query the capabilities of the installed switching modules.
For example, the following queries are provided to determine which channels of a
Model 7700 in slot 1 are available for volts/2-wire ohms and which channels are available
for amps. Note that the returned values for the Model 7700 switching module are shown in
parenthesis.
SYSTem:CARD1:VCHannel[:STARt]? ' Query first volt/2 channel(7700; "1").
SYSTem:CARD1:VCHannel:END? ' Query last volt/2 channel (7700; "20").
SYSTem:CARD1:ACHannel[:STARt]? ' Query first amps channel (7700; "21").
SYSTem:CARD1:ACHannel:END? ' Query last amps channel (7700; "22").
All the commands to query switching module capabilities are covered in Table 15-7.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals