Model 2701 User’s Manual SCPI Signal Oriented Commands 13-3
NOTE The CONFigure:<function> and MEASure:<function>? commands can be sent
without any of the optional parameters (<rang>, <res>, <clist>). For details,
see the “Description” for the CONFigure and MEASure commands.
When using the <clist> parameter, it is interpreted as the last parameter. Any
parameter after <clist> will generate error -102 (syntax error).
If only one parameter is used and it is not a <clist>, it is interpreted as the
range parameter (<rang>).
If two parameters are used and the second one is not a <clist>, the first
parameter is the range parameter (<rang>) and the second is the resolution
parameter (<res>).
The CONFiguration and MEASure? commands for the TEMPerature and
CONTinuity functions do not use the <rang> and <res> parameters. The
command is ignored and causes error -108 (parameter not allowed).
The CONFigure and MEASure? commands cannot be used while scanning. The
command is ignored and causes error -221 (settings conflict).