4-18 Range, Digits, Rate, Bandwidth, and Filter Model 2701 User’s Manual
Filter control and conguration
The FILTER key toggles the state of the Filter. When the Filter is enabled, the FILT
annunciator is on. The FILT annunciator will flash when the filter is not settled. When
disabled, the FILT annunciator is off. The filter can be configured while it is enabled or
disabled.
The filter is configured from the filter configuration menu (Figure 4-4). Perform the
following steps to configure the filter:
1. Select the desired function.
2. Press SHIFT and then TYPE. The present WINDOW setting will be displayed.
3. Use the RANGE
or key to display the desired window setting (0.01%, 0.1%,
1%, 10%, or NONE) and press ENTER.
4. Use the , , , and keys to display the number of readings to filter (1 to 100)
and press ENTER.
5. Use the or key to display the desired filter type (moving or repeating) and
press ENTER. The filter turns on and the instrument returns to the normal
measurement state.
NOTE While the filter is enabled (FILT annunciator on), changes to the configuration
take effect as soon as they are made. With filter disabled (FILT annunciator off),
changes to the configuration take place when the filter is enabled.
While the filtering operation is in progress, the FILT annunciator blinks. Read-
ings will continue to be processed (i.e., displayed, stored, sent over the bus), but
they could be questionable. When the FILT annunciator stops blinking, the filter
has settled.
Changing function or range causes the filter to reset. The filter then assumes the
state (enabled or disabled) and configuration for that function or range.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals