Model 2701 User’s Manual Rel, Math, Ratio, Channel Average, dB 5-5
Commands
1
Description Default
Rel commands for
2
[SENSe[1]] Optional root command.
:RESistance:REFerence <n> [, <clist>] Specify rel value; <n> = 0 to 120e6 (). 0
:RESistance:REFerence:STATe <b> [, <clist>] Enable/disable rel; <b> = ON or OFF. OFF
:RESistance:REFerence:ACQuire [, <clist>] Use input signal as rel value.
Rel commands for
4
[SENSe[1]] Optional root command.
:FRESistance:REFerence <n> [, <clist>] Specify rel value; <n> = 0 to 120e6 (). 0
:FRESistance:REFerence:STATe <b> [, <clist>] Enable/disable rel; <b> = ON or OFF. OFF
:FRESistance:REFerence:ACQuire [, <clist>] Use input signal as rel value.
Rel commands for
TEMP
2
[SENSe[1]] Optional root command.
:TEMPerature:REFerence <n> [, <clist>] Specify rel value; <n> = -328 to 3310 (°C). 0
:TEMPerature:REFerence:STATe <b>
[, <clist>]
Enable/disable rel; <b> = ON or OFF. OFF
:TEMPerature:REFerence:ACQuire [, <clist>] Use input signal as rel value.
Rel commands for FREQ
[SENSe[1]] Optional root command.
:FREQuency:REFerence <n> [, <clist>] Specify rel value; <n> = 0 to 1.5e7 (Hz). 0
:FREQuency:REFerence:STATe <b> [, <clist>] Enable/disable rel; <b> = ON or OFF. OFF
:FREQuency:REFerence:ACQuire [, <clist>] Use input signal as rel value.
Rel commands for PERIOD
[SENSe[1]] Optional root command.
:PERiod:REFerence <n> [, <clist>] Specify rel value; <n> = 0 to 1 (sec). 0
:PERiod:REFerence:STATe <b> [, <clist>] Enable/disable rel; <b> = ON or OFF. OFF
:PERiod:REFerence:ACQuire [, <clist>] Use input signal as rel value.
Channel list parameter:
<clist> = (@SCH)
where: S = Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
Examples: (@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
1. The <clist> parameter is used to configure one or more channels for a scan. Each channel in the <clist> must be set to the function
specified by the rel (reference) command. If not, a conflict error (-221) will occur. For example, VOLTage:AC:REFerence 1,
(@101) is only valid if scan channel 101 is set for the ACV function.
2. [:DC] is optional for the commands to set DCV and DCI rel.
Table 5-1 (continued)
Rel commands

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm