5-14 Rel, Math, Ratio, Channel Average, dB Model 2701 User’s Manual
Math programming examples
Example #1 — The following command sequence performs the mX+b calculation for
channels 101 and 102 of the Model 7700.
NOTE The following example can be run from the KE2700 Instrument Driver using the
example named “Linear” in Table G-1 of Appendix G.
CALC:FORM MXB Select mX+b calculation.
CALC:KMAT:MMF 2 Set m factor to 2.
CALC:KMAT:MBF 100 Set b factor to 100.
CALC:STAT ON Enable math calculation.
ROUT:CLOS (@101) Close channel 101.
CALC:DATA? Read mX+b result for channel 101.
ROUT:CLOS (@102) Close channel 102.
CALC:DATA? Read mX+b result for channel 102.
Example #2 — The following command sequence configures channels 101 through 110
of the Model 7700 to perform the percent calculation when they are scanned.
NOTE The following example can be run from the KE2700 Instrument Driver using the
example named “Percent” in Table G-1 of Appendix G.
CALC:FORM PERC,(@101:110) Select percent calculation.
CALC:KMAT:PERC 100,(@101:110) Set reference to 100.
CALC:STAT ON,(@101:110) Enable math calculation.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals