Model 2701 User’s Manual Scanning 7-17
Step 3: Enable immediate scan
The present state of immediate scan (IMM SCAN) is displayed, Y (yes, which is the
factory and *RST default) or N (no). With immediate scan enabled, the scan will start
when you press the STEP or SCAN key. Use the or key to display IMM SCAN: Y and
press ENTER.
NOTE Disable immediate scan (IMM SCAN: N) when you wish to use a monitored
reading limit to trigger the start of the scan. This technique to start a scan is
covered later in this section. See “Scan operation,” page 7-21.
Step 4: Timer controlled scan
The present state of the TIMER will be displayed, NO or YES. If you do not wish to use
the timer, use the
or key to display TIMER? NO (which is the factory and *RST
default) and press ENTER.
To use the timer, use the or key to display TIMER? YES and press ENTER. The timer
interval is displayed in the hour/minute/second format. The timer can be set from 0.001
sec (00H:00M:00.001S) to 99 hr, 99 min, 99.999 sec (99H:99M:99.999S). Note that
pressing the AUTO key sets the timer to 0.001 sec. With the desired interval displayed,
press ENTER.
Step 5: Set the reading count
The displayed reading count (RDG CNT) sets the number channels to scan (STEP) or the
number of scans to run (SCAN). You can change the reading count to any finite value from
2 to 450000 or you can select infinite (continuous scanning). To select infinite, set the
reading count to 00000 to display INF. With the desired reading count setting displayed,
press ENTER. The instrument returns to the normal measurement mode.
See “Trigger models,” page 7-4, for details on reading count.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm