Model 2701 User’s Manual Scanning 7-25
2. Press the or key to display IMM SCAN: N and press ENTER.
a. Press the or key to enable or disable low limit 1 (LLIM1 SCAN:N/Y) and
press ENTER.
b. Press the or key to enable or disable high limit 1 (HLIM1 SCAN:N/Y)
and press ENTER.
c. Press the or key to enable or disable low limit 2 (LLIM2 SCAN:N/Y) and
press ENTER.
d. Press the or key to enable or disable high limit 2 (HLIM2 SCAN:N/Y)
and press ENTER.
3. Finish configuring the scan by performing Step 4 and Step 5 of the Advanced scan
setup procedure,” page 7-16.
NOTE For a remote programmed monitor scan, use the ROUTe:MONitor:POINts
command to specify the number of channels to scan (Table 7-1).
4. While in the normal measurement state, select and enable the monitor channel as
explained in “Scan configurationMonitor channel,” page 7-18.
When the reading limit for the monitor channel is reached, the scan will be
triggered to start. When the monitor channel is scanned, the display will show the
reading that triggered the scan.
If the reading limit event is still present on the monitor channel when the scan
finishes, the scan will be triggered to run again. Note that the scan can also be run
by pressing STEP or SCAN.
5. To disable the monitor scan, perform the following steps:
a. To disable monitor, press SHIFT and then MONITOR (MON annunciator
turns off).
b. To disable limits, press SHIFT and then ON/OFF. Press
or to display
“LIMITS: OFF” and press ENTER.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals