Model 2701 User’s Manual Remote Operations 10-17
Local intranet address
Using Internet Explorer, a local intranet address is used to open the web page. The web
page address is the same as the IP Address of the Model 2701. Assuming the default IP
Address is being used, the following intranet address will open the web page:
http://192.168.0.2
If the use of password is enabled, you will be prompted to enter the password. If you for-
got the password, you can either contact your network administrator for the password or
you can clear the password from the front panel of the Model 2701. To clear the password
and open the web page, perform the following:
1. Model 2701:
Press LOCAL > press SHIFT > press SETUP > display
CLR PASSWORD:Y > press ENTER
Clearing the password disables its use and resets it to “DEFAULT”.
2. PC:
Click Cancel on the Login dialog box.
Click Refresh on the IE tool bar to open the web page.
With the web page open, the use of password can be enabled and changed (if
desired).
Diagnostic summary
The diagnostic summary, shown in Figure 10-9, provides the various details about the
Model 2701. The fields with the grey background are read-only details and cannot be
changed. Fields with white background can be changed. As shown in Figure 10-9, you can
either click the Accept Changes button to make the changes or Restore Details to cancel
the changes.
NOTE If you change the Ethernet settings and they conflict with the PC, you will lose
connection to the Model 2701. Ethernet settings for the Model 2701 can be fixed
from the front panel.
Interactive control panel
The control panel allows the user to control the Model 2701. You can send commands and
queries. Readings, responses to queries, and error messages are displayed in the control
panel. Details on using the control panel are provided in Figure 10-10.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm