10-26 Remote Operations Model 2701 User’s Manual
Command path rules
Each new program message must begin with the root command, unless it is
optional (e.g., [SENSe]). If the root is optional, simply treat a command word on
the next level as the root. For fastest operation, do not send optional data.
A colon (:) can be used at the beginning of a program message. However, using the
colon slows down execution time. Example:
stat:pres = :stat:pres
When the path pointer detects a colon (:) it moves down to the next command level.
An exception is when the path pointer detects a semicolon (;), which is used to
separate commands within the program message (see next rule).
When the path pointer detects a colon (:) that immediately follows a semicolon (;),
it resets back to the root level.
The path pointer can only move down. It cannot move up a level. Executing a
command at a higher level requires that you start over at the root command.
Using common and SCPI commands in the same message
Both common commands and SCPI commands can be used in the same message as long
as they are separated by semicolons (;). A common command can be executed at any
command level and will not affect the path pointer. Example:
stat:oper:enab <NRf>; *ESE <NRf>
Program message terminator (PMT)
Each program message must be terminated with an LF (line feed), CR (carriage return),
CR-LF or LF-CR. The command processor will hang if your computer does not provide
this termination. The following example shows how a multiple command program
message must be terminated:
trig:coun 1 <PMT>
Command execution rules
Commands execute in the order that they are presented in the program message.
An invalid command generates an error and, of course, is not executed.
Valid commands that precede an invalid command in a multiple command program
message are executed.
Valid commands that follow an invalid command in a multiple command program
message are ignored.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm