11-12 Status Structure Model 2701 User’s Manual
Bit B3, Device-Dependent Error (DDE) — Set bit indicates that an instrument
operation did not execute properly. Some of the errors specific to the Model 2701
that will set this bit include the following:
Error +516: Battery backed RAM error — Data stored in RAM has been lost.
Replace the battery if frequent failures occur.
Error +517: Cannot resume scan — Due to a card ID change, auto scan has
disabled. The scan will not resume after a power interruption. For details, see
Scan configurationAuto scan,” page 7-21.”
Error +520: Saved setup scancard mismatch — Settings for a user setup or
power-on setup do not match the switching module types presently installed.
For details, see “Defaults and user setups,” page 1-22.
Error +523: Card hardware error — Communication with the microprocessor
on a switching module card has been lost.
Error +524: Unsupported card detected — The Model 2701 has detected an
installed Model 77XX switching module that is not supported by the current
version of firmware.
Bit B4, Execution Error (EXE) — Set bit indicates that the Model 2701 detected
an error while trying to execute a command.
Bit B5, Command Error (CME) — Set bit indicates that a command error has
occurred:
IEEE-488.2 syntax error – Instrument received a message that does not follow
the defined syntax of the IEEE-488.2 standard.
Semantic error – Instrument received a command that was misspelled or
received an optional IEEE-488.2 command that was not implemented.
The instrument received a Group Execute Trigger (GET) inside a program
message.
Bit B6, User Request (URQ) — Set bit indicates that the LOCAL key on the
Model 2701 front panel was pressed.
Bit B7, Power On (PON) — Set bit indicates that the Model 2701 has been turned
off and turned back on since the last time this register has been read.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals