11-14 Status Structure Model 2701 User’s Manual
Measurement event register
The used bits of the Measurement Event Register (Figure 11-6) are described as follows:
Bit B0, Reading Overflow (ROF) — Set bit indicates that the reading exceeds the
measurement range of the instrument.
Bit B1, Low Limit 1 Event (LL1) — Set bit indicates that a reading has reached or
exceeded Low Limit 1.
Bit B2, High Limit 1 Event (HL1) — Set bit indicates that a reading has reached
or exceeded High Limit 1.
Bit B3, Low Limit 2 Event (LL2) — Set bit indicates that a reading has reached or
exceeded Low Limit 2.
Bit B4, High Limit 2 Event (HL2) — Set bit indicates that a reading has reached
or exceeded High Limit 2.
Bit B5, Reading Available (RAV) — Set bit indicates that a reading was taken and
processed.
NOTE A programming example to read the RAV bit is provided in “Example 2 – Read
RAV bit of measurement event register” on page 11-20.
Bit B6, Buffer Notify (BN) — Set bit is a notification that the user-specified
number of readings have been stored in the buffer. The TRACe:NOTify command
specifies the number of stored readings that will set this bit (see Section 6 for
details).
Bit B7, Buffer Available (BAV) — Set bit indicates that there are at least two
readings in the buffer.
Bit B8, Buffer Half Full (BHF) — Set bit indicates that the trace buffer is half
full.
NOTE A programming example to read the BHF bit is provided in “Example 3 – Read
BHF bit of measurement event register” on page 11-20.
Bit B9, Buffer Full (BF) — Set bit indicates that the trace buffer is full.
Bit B10, Buffer Overflow (BOF) — Set bit indicates that the filled buffer has
wrapped and written over previously stored readings.
Bit B11, Hardware Limit Event (HL) — Set bit indicates that a reading has
exceeded the hardware limit.
Bit B12, Buffer Quarter Full (BQF) — Set bit indicates that the trace buffer is
one-quarter full.
NOTE Bits B12 (¼ full) and B13 (¾ full) are not intended to be used with buffer sizes
smaller than four readings.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm