Model 2701 User’s Manual Common Commands 12-5
Programming example – The following command sequence demonstrates how to use
*OPC? to signal the end of a measurement process:
SYST:PRES ' Returns 2701 to default setup.
INIT:CONT OFF ' Disables continuous initiation.
ABORt ' Aborts operation. Places 2701 in idle.
TRIG:COUN 1 ' These two commands configure the 2701
SAMP:COUN 5 ' to perform five measurements.
INIT ' Starts measurement process.
*OPC? ' Sends OPC? command.
After all five measurements are performed and the instrument returns to the idle state, an
ASCII ‘1’ will be placed in the Output Queue. The ‘1’ from the Output Queue is sent to
the computer.
SYST:PRES ' Returns 2701 to default setup.
NOTE The following commands take a long time to process and may benefit from using
*OPC or *OPC?:
*RST and SYST:PRES
*RCL and *SAV
ROUT:MULT:CLOS and ROUT:MULT:OPEN – Only if the <clist> is long.
CALC2:IMM and CALC2:IMM?– Only when performing the standard
deviation calculation on a large buffer. A 450,000 point buffer takes around
5 to 6 seconds.
D *OPT? option query Query installed switching modules
Use this query command to determine which switching modules are installed in the
Model 2701. For example, if a Model 7703 is installed in slot 1 and the other slot is empty,
the response message will look like this:
7703, NONE
Note that the model number of an installed pseudocard is returned in the same manner.
See Section 2 for details on pseudocards.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
Max Voltage300 V
Max Current3 A
Frequency Range3 Hz to 500 kHz
Frequency Accuracy0.01%
Weight3.6 kg
Measurement FunctionsFrequency, Temperature
InterfaceIEEE-488 (GPIB), RS-232
DC Voltage Range300 V
DC Current Range3 A
AC Voltage Range300V
AC Current Range1 μA to 3 A
Resistance Accuracy0.01%
Channels10
Dimensions213mm x 88mm x 370mm

Related product manuals