Model 2701 User’s Manual Getting Started 1-19
The instrument then briefly displays all display characters and annunciators, and performs
self-tests. If all self-tests pass, the firmware revision levels are displayed. An example of
this display is:
REV: A01 A01
where: First A01 is the main board ROM revision.
Second A01 is the display board ROM revision.
If Ethernet is the presently selected interface, the IP address will be displayed briefly.
Example:
ETHERNET
192.168.000.002
If RS-232 is the presently selected interface, the baud rate will be displayed briefly:
RS-232 MODE
BAUD: 9600
Installed switching modules are then displayed. For example, if there is a Model 7700
switching module installed in both slots, the following messages will be displayed:
1: 7700 2: 7700
If a slot is empty, the message “NONE” will be displayed instead.
If the saved power-on setup is not the factory defaults setup (SYSTem:POSetup PRESet),
a message to identify the setup will be briefly displayed (“Defaults and user setups,”
page 1-22).
After the power-up sequence, the instrument begins its normal display of readings.
NOTE The serial number of the Model 2701 can be displayed by selecting the SNUM
item of the SETUP menu. Press SHIFT and then SETUP to access the menu. For
remote operation, the serial number can be read using the *IDN? command (see
Section 12 for details).
NOTE The serial number and revision levels can be accessed from the internal web
page of the Model 2701 using the Internet Explorer. For details, see “Internal
web page” in Section 10.