D-4 Signal Processing Sequence and Data Flow Model 2701 User’s Manual
OComp (offset-compensated ohms)
The Model 2701 performs a normal ohms measurement by sourcing a known current (I),
measuring the voltage (V), and then calculating the resistance (R = V/I). Offset-
compensated ohms cancels the effects of thermal EMFs which can adversely affect low-
resistance measurements.
With OCOMP enabled, the Model 2701 performs one normal resistance measurement and
then loops back to perform a second resistance measurement with the internal current
source set to its lowest level. The offset-compensated ohms reading is then calculated as
shown in Figure D-2.
NOTE For details on OCOMP measurements, see“Offset-compensated ohms,
page 3-24.
Filter
The filter is used to stabilize noisy readings. With the filter enabled, the specified number
of readings are averaged to yield a single filtered reading. There are two types of filters:
moving and repeating.
A filter stack is used to temporarily store the specified number of readings to be averaged.
In general, for the moving filter, each measurement process adds a reading to the stack
(oldest reading discarded) and then averages the stack to yield a filtered reading. For the
repeating filter, each measurement process fills the stack with new readings (all previous
readings discarded) and then averages the stack to yield a filtered reading.
NOTE For details on filter operation, see “Filter,” page 4-14.
Output trigger pulse (VMC)
An output trigger pulse from the Model 2701 can be used to trigger an external instrument
to perform an operation. In general, a trigger pulse is the output at this point in the
flowchart for each processed reading.
An exception is the SCAN function for scanning. For the SCAN function, an output
trigger is not output until after the specified number of channels (as set by the sample
counter) are scanned.
NOTE For details on scan operation, see “Trigger models,” page 7-4.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals