EasyManuals Logo
Home>Keithley>Multimeter>2701

Keithley 2701 User Manual

Keithley 2701
506 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #468 background imageLoading...
Page #468 background image
F-4 Temperature Equations Model 2701 User’s Manual
Table F-4
Type K inverse function polynomial
-200°C to 0°C
(-5,891µV to 0µV)
0°C to 500°C
(0µV to 20,644µV)
500°C to 1,372°C
(20,644µV to 54,886µV)
c
0
= 0.0 0.0 -1.318 058 × 10
2
c
1
= 2.517 346 2 × 10
-2
2.508 355 2 × 10
-2
4.830 222 × 10
-2
c
2
= -1.166 287 8 × 10
-6
7.860 106 2 × 10
-8
-1.646 031 ×10
-6
c
3
= -1.083 363 8 × 10
-9
-2.503 131 2 × 10
-10
5.464 731 × 10
-11
c
4
= -8.977 354 0 × 10
-13
8.315 270 2 × 10
-14
-9.650 715 × 10
-16
c
5
= -3.734 237 7 × 10
-16
-1.228 034 2 × 10
-17
8.802 193 × 10
-21
c
6
= -8.663 264 3 × 10
-20
9.804 036 2 × 10
-22
-3.110 810 × 10
-26
c
7
= -1.045 059 8 × 10
-23
-4.413 030 2 × 10
-26
c
8
= -5.192 057 7 × 10
-28
1.057 734 2 × 10
-30
c
9
= -1.052 755 2 × 10
-35
Error: 0.04°C to -0.02°C 0.04°C to -0.05°C 0.06°C to -0.05°C
t
90
= c
0
+ c
1
E + c
2
E
2
+ c
3
E
3
... c
i
E
i
where: t
90
is the calculated temperature in °C. E is the measured voltage in microvolts.
Table F-5
Type N inverse function polynomial
-200°C to 0°C
(-3,990µV to 0µV)
0°C to 600°C
(0uV to 20,613µV)
600°C to 1,300°C
(20,613µV to 47,513µV)
c
0
= 0.0 0.0 1.972 485 × 10
1
c
1
= 3.843 684 7 × 10
-2
3.868 96 × 10
-2
3.300 943 × 10
-2
c
2
= 1.101 048 5 × 10
-6
-1.082 67 × 10
-6
-3.915 159 × 10
-7
c
3
= 5.222 931 2 × 10
-9
4.702 05 × 10
-11
9.855 391 × 10
-12
c
4
= 7.206 052 5 × 10
-12
-2.121 69 × 10
-18
-1.274 371 × 10
-16
c
5
= 5.848 858 6 × 10
-15
-1.172 72 × 10
-19
7.767 022 × 10
-22
c
6
= 2.775 491 6 × 10
-18
5.392 80 × 10
-24
c
7
= 7.707 516 6 × 10
-22
-7.981 56 × 10
-29
c
8
= 1.158 266 5 × 10
-25
c
9
= 7.313 886 8 × 10
-30
Error: 0.03°C to -0.02°C 0.03°C to -0.02°C 0.02°C to -0.04°C
t
90
= c
0
+ c
1
E + c
2
E
2
+ c
3
E
3
... c
i
E
i
where: t
90
is the calculated temperature in °C. E is the measured voltage in microvolts.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals