Service 2
Keysight 53210A/53220A/53230A Assembly Level Service Guide 173
Re-run the counter self test
Even if you ran the self-test previously, after reseating the boards the self-test
should be run again. (Refer to page 169 .)
NOTE
After diagnosing an error or problem and isolating the failure to an assembly,
refer to Chapter 3, "Replacing Assemblies" in this service guide for assembly
removal and replacement procedures.
NOTE
After replacing an assembly, perform the calibration procedures in this chapter.
CAUTION
DO NOT swap the motherboard, the processor board, or the front panel
board from one instrument to another. These boards contain model number
and serial number information that uniquely identifies a specific unit, and
boards that are mismatched to the instrument may result in problems with
its performance, licensing, serviceability, importability/exportability or
warranty. It is also likely that unpredictable error messages could occur.
(See “Error Messages” on page 174.)

Table of Contents

Other manuals for Keysight Technologies 53210A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 53210A and is the answer not in the manual?

Keysight Technologies 53210A Specifications

General IconGeneral
BrandKeysight Technologies
Model53210A
CategoryCash Counter
LanguageEnglish

Related product manuals