Replacing Assemblies 3
Keysight 53210A/53220A/53230A Assembly Level Service Guide 183
Tools Required
The following tools are required for these removal and replacement procedures:
–Hand TORX 8 screwdriver (T8)
Hand TORX‚ 10 screwdriver (T10)
Hand TORX‚ 20 screwdriver (T20)
7-mm spin tight
14-mm deep-socket spin tight
–1/4" spin tight
5/16" open end wrench
Static-Safe Workstation
Do this first
Before performing any of the removal and replacement procedures, the following
steps must be performed.
1 Turn counter off.
2 Remove AC line power cord from the rear AC line socket.
3 Open up the counter only at a static-safe workstation.
WARNING
Hazardous voltages are on the power supply assembly. After unplugging the
instrument, wait at least 6 minutes for the capacitors on the power supply
assembly to discharge before working on the instrument.