EasyManua.ls Logo

Keysight Technologies 53210A - Test 7 c: Pulse Burst Frequency (1.35 GHz)

Keysight Technologies 53210A
273 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests 1
Keysight 53210A/53220A/53230A Assembly Level Service Guide 83
Test 7c: (Q) Pulse burst frequency - 53230A channel 3 option 150 with
option 115
This test uses the 53230A to measure both narrow and wide pulse outputs of a
1.35 GHz carrier using the Opt 115 microwave channel.
Equipment
FS725 Rubidium timebase
Keysight E8257D-520-UNW microwave signal generator
Keysight 10503A BNC cable (2)
Keysight 11500E (3.5) mm cable
Keysight 1250-1250 N(M) to SMA(F)
E8257D-520-UNW setup
Output Frequency: 1.35 GHz
Output Power: -6 dBm
–Auto-Level Control: Off
–Pulse Mode: On
Pulse Repetition Interval Period: 1 mSec
Pulse Width (Narrow): 400 nSec
NOTE
This test procedure applies only to the 53230A Counter with Option 115 and
Option 150 installed. Option 115 is installed as microwave channel 3.
NOTE
The Quick Verification (Q) requires only the Narrow Pulse Width Burst Frequency
Test be performed.

Table of Contents

Other manuals for Keysight Technologies 53210A

Related product manuals